Euglobulin lysis time
(ELT)
Sample type:
blood sample, plasma
Uses:
a test measures overall fibrinolysis.
precautions
• When drawing the sample, be careful not to rub the area over the vein too vigorously, pump the fist excessively, or leave the tourniquet in place too long.
• Avoid excessive probing during venipuncture, and handle the sample gently.
• If a blue-top tube is used, mix the sample and anticoagulant thoroughly. If a chilled tube containing 0.5 ml sodium oxalate is used, mix the sample and preservative thoroughly, pack the sample in ice, and send it to the laboratory immediately.
Interfering factor:
• Prolonged tourniquet constriction, vigorous vein preparation, or excessive pumping of the fist (decrease).
• Hemolysis due to excessive probing at the venipuncture site or to rough handling of the sample.
• Failure to place the collection tube and sample on ice
• Fibrinogen level < 100 mg/dl (possible decrease)
Pre-analytical error:
• Coagulation samples should not be transported or stored on ice
• Coagulation samples for platelet function studies must be maintained at room temperature
• Gentle inversion (mixing) of sodium citrates tube 5–6 times is recommended. Avoid rigorous shaking or agitation 69 3.2% sodium citrate is the citrate concentration of choice
• If the syringe technique is to be used, then a syringe of less than 25 mL (preferably 10 mL) using a “butterfly” needle apparatus
Post analytical errors:
• Check instrument settings
• Assure that all preventive maintenance has been done
• Carry out troubleshooting functions called for in the SOP
• Call for equipment support
Corrective action:
• Assure that the prescribed preventive maintenance for the instrument is performed
Reference Ranges
Hyperfibrinolysis: <60 minutes
Hypofibrinolysis: >240 minutes